Correct use method of liquid nitrogen cryogenic box
2020-12-23 09:12:48
What is the correct way to use the liquid nitrogen cryogenic box?
Liquid nitrogen cryogenic box is a kind of liquid nitrogen cryogenic equipment, which is widely used in low temperature treatment, heat treatment, cold assembly of metal materials or I-beam, food quick freezing industry, electronic product touch screen separation, scientific experiment and other fields. Correct use and operation The equipment not only helps to extend the service life of the equipment, but also helps to save costs. So, how can the device be operated correctly?
The detailed operation process of liquid nitrogen cryogenic equipment is as follows:
One, turn on
1. Turn on the main switch of the power distribution box for the equipment;
2. Turn on the equipment leakage power switch (on the left side of the equipment), the normal equipment controller will start to light up and display the operation screen, if not, first check the emergency stop switch, turn on the emergency stop switch in the indicated direction; contact if you have any questions factory.
2. Enter the operation interface
Start the device to enter the controller operation interface, set the required test temperature, click to run the device to start the test. Note that the door lock is locked and sealed well. Enter the operation interface of the liquid nitrogen cryogenic box and click "Detail" to enter the following section, and more test information will be displayed.
Three, start operation
If the cryogenic refrigeration system needs to be tested from -120°C to -160°C, liquid nitrogen cooling is required, and the liquid nitrogen valve needs to be opened manually. The operation is as follows:
1. Check the pressure of the liquid nitrogen dewar. When the pressure is lower than 7 kg, open the booster valve and slowly open it. When the pressure gauge indicates more than 10 kg, close the booster valve.
2. After checking the previous work, start the test and set the required test temperature (below -120℃ to -160℃);
3. When the controller shows that the temperature reaches -120℃, the liquid nitrogen needs to be turned on for the lower temperature test. At this time, slowly open the liquid nitrogen valve, use a small opening to avoid excessive cooling of the equipment, and observe the cooling situation (the equipment starts Continue to decrease temperature)
4. Don't close the valve after the temperature has dropped. The equipment is controlled by solenoid valve. After the experiment is over, close the liquid nitrogen valve.
Be careful not to open the door during the test. When opening the door in an emergency, pay attention to prevent freezing, and people should not be directly exposed to the air-conditioning.
Four, turn off
1. Click "Stop" for controller operation;
2. Perform oven operation: set the temperature to 60°C;
3. When the temperature of the liquid nitrogen cryogenic box reaches 60℃, it can be baked for about 10-20 minutes to avoid excessive condensation in the box and the evaporator;
4. Turn off the main power switch.
Nuoruida (Tianjin) science and Technology Co., Ltd. is committed to research, development, and sales of liquid nitrogen freezer. Using European technology, we provide a variety of equipment. The main equipment is liquid nitrogen Cryogenic treatment equipment, cryogenic freezer, liquid nitrogen cabinet freezer, liquid nitrogen tunnel freezer,sub-zero processing equipment, low temperature cold assembly equipment, cryogenic freezing equipment, and so on.